【Summary】 The Impact of Temperature on Measurement Size We conduct a large number of various measurements every day, striving for measurement accuracy during the process. Today, we will briefly discuss the impact of temperature on length measurement. 1. Factors Affecting Temperature Error In measurement conditions, factors such as temperature, humidity, vibration, dust, and corrosive gases can directly or indirectly affect measurement accuracy. Among these factors, the variation in temperature has a particularly significant impact on measurement accuracy. Due to the large differences in the expansion coefficients of different materials, the principle of equalizing the temperature of the measured object with that of the measuring standard instrument is adopted during measurement. Various measurement professions should work according to the temperature conditions required by the verification procedures to ensure the accuracy of the detection data. Currently, the standard temperature for measurement rooms in our country is set at 20℃. By consulting the manual, the relationship between the linear expansion coefficient a, temperature change, and size change can be expressed as follows: In the formula: a is the linear expansion coefficient; △L is the size change; L is the object size; △T is the temperature change. When a workpiece experiences size changes due to its temperature deviating from 20℃, it can be expressed as: In the formula: t is the temperature of the object. When both the workpiece and the measuring tool have deviations from the standard temperature, the measurement error caused by temperature is the difference in size changes of the two, which can be expressed as: In the formula: △L is the size change; L is the object size; a1 and a2 are the linear expansion coefficients of the workpiece and measuring tool materials; t1 and t2 are the temperatures of the workpiece and measuring tool. Definition of the linear expansion coefficient of an object: The size change of a unit length (1mm) when the temperature changes by 1℃ is only related to the material of the object. Example: Suppose a micrometer measures a copper shaft with a diameter of ∅100mm, where the shaft's temperature is 40℃ and the micrometer's temperature is 15℃. The linear expansion coefficients of the copper shaft and the steel micrometer are 17.5 x10-6 and 11.5 x10-6, respectively. The size change △L is: It can be seen that temperature has a significant impact on the size of the workpiece, with this example showing a difference of several tens. Therefore, when measuring measuring instruments, it is necessary to measure after the workpiece and measuring tool have reached thermal equilibrium, which will greatly reduce temperature errors. The time required for temperature balancing during the verification of each measuring tool is specified in the corresponding verification procedures and calibration specifications. Note: The process in which the temperature of the measured tool and the standard tool tends to be consistent under certain conditions (such as at room temperature of (20±6)℃) is called thermal equilibrium. 2. Methods to Reduce Temperature Errors 1. Measure when the temperature is close to the standard temperature. In actual work, we also strictly follow relevant procedures, ensuring that the measuring instruments sent by customers reach thermal equilibrium with the measurement room temperature before verification or calibration. Therefore, if customers want to obtain measurement data immediately, the data will definitely be affected by temperature. 2. When measuring on-site, do not take out the standard instrument and measure immediately. Instead, place the standard instrument and the workpiece together, and let them reach thermal equilibrium on a large flat surface before measuring. 3. Avoid the influence of hand temperature on the workpiece and measuring tool. For example, when using and verifying the micrometer, it is necessary to hold it with an insulating pad. When it is found that the heat from the hand has been transferred to the micrometer, it should be placed aside for a certain period before use and verification. 4. When measuring large sizes outdoors, it is necessary to use measuring instruments for temperature compensation according to the corresponding materials to minimize temperature errors.
The effect of temperature on measurement size
Every day we conduct a large number of various measurements, pursuing accuracy in the measurement process. Today, we will briefly discuss the impact of temperature on length size measurement.
1. Factors Affecting Temperature Error
In measurement conditions, factors such as temperature, humidity, vibration, dust, and corrosive gases can directly or indirectly affect measurement accuracy. Among these factors, the variation in temperature has a particularly significant impact on measurement accuracy. Due to the large differences in the expansion coefficients of different objects, the principle of using the temperature of the measured object equal to that of the measuring standard instrument should be adopted during measurement. Various measurement professions should work according to the temperature conditions required by the verification procedures to ensure the accuracy of the detection data. Currently, the standard temperature for measurement rooms in our country is set at 20℃.
According to the manual, the relationship between the linear expansion coefficient a and the changes in temperature and size can be expressed as follows, represented by formula (1):
In the formula: a is the linear expansion coefficient; △L is the size change; L is the object size; △T is the temperature change.
The size change caused by a workpiece due to its temperature deviation from 20℃ can be expressed by formula (2):
In the formula: t is the object temperature.
When both the workpiece and the measuring tool have deviations from the standard temperature, the measurement error caused by temperature is the difference in size changes of the two, which can be expressed by formula (3):
In the formula: △L is the size change; L is the object size; a1 and a2 are the linear expansion coefficients of the workpiece and the measuring tool materials; t1 and t2 are the temperatures of the workpiece and the measuring tool.
Definition of the linear expansion coefficient of an object: the size change of a unit length (1mm) when the temperature changes by 1℃, which is only related to the material of the object.
Example: Suppose a micrometer is used to measure a copper shaft with a diameter of ∅100mm, the temperature of the shaft is 40℃, and the external micrometer is at 15℃. The linear expansion coefficients of the copper shaft and the steel external micrometer are 17.5 x10-6 and 11.5 x10-6, respectively, then the size change △L is:
It can be seen that temperature has a significant impact on the size of the workpiece, with this example showing a difference of several tens of units. Therefore, when measuring measuring instruments, it is necessary to measure after the workpiece and the measuring tool have reached thermal equilibrium, which will greatly reduce the temperature error. The time required for thermal equilibrium for each measuring tool is specified in the corresponding verification procedures and calibration specifications.
Note: The process in which the measured tool and the standard tool tend to have the same temperature under certain conditions (such as at room temperature of (20±6)℃) is called thermal equilibrium.
2. Methods to Reduce Temperature Error
1. Measure when the temperature is close to the standard temperature. In actual work, we also strictly follow the relevant procedures, allowing the measuring instruments sent by customers to reach thermal equilibrium with the measurement room temperature before verification or calibration. Therefore, if customers want to obtain measurement data immediately, the measurement data will definitely be affected by temperature.
2. When measuring on-site, do not take out the standard instrument and measure immediately. Instead, place the standard instrument and the workpiece together and allow them to reach thermal equilibrium on a large flat surface before measuring.
3. Avoid the influence of hand temperature on the workpiece and measuring tool. For example, when using and verifying the micrometer, it is necessary to hold the insulating pad. When it is found that the heat from the hand has been transferred to the micrometer, it should be placed for a certain time before use and verification.
4. When measuring large sizes outdoors, it is necessary to use measuring instruments for temperature compensation according to the corresponding materials to minimize temperature errors.